Appro LC-7451 Instruções de Operação

Consulte online ou descarregue Instruções de Operação para Câmaras de segurança Appro LC-7451. Appro LC-7451 Operating instructions Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
Thermo Scientific Dionex
UltiMate 3000 Series
Variable Wavelength Detectors
VWD-3100 and VWD-3400RS
Operating Instructions
(Original Operating Instructions)
Revision: 1.7
Date: September 2013
© 2013 Thermo Fisher Scientific Inc.
Doc No. 4820.7451
Vista de página 0
1 2 3 4 5 6 ... 159 160

Resumo do Conteúdo

Página 1 - UltiMate 3000 Series

Thermo Scientific Dionex UltiMate 3000 Series Variable Wavelength Detectors VWD-3100 and VWD-3400RS Operating Instructions (Original Operating

Página 2 - Operating Instructions

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 2 Operating Instructions All rights reserved, including those for

Página 3 - Declaration of Conformity

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 92 Operating Instructions 5.7 Shutting Down the Detector Observe

Página 4

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 93 To create a standby or shutdown program

Página 5 - Table of Content

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 94 Operating Instructions 5.8 Routine and Preventive Maintenance

Página 6

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 95 6 Troubleshooting 6.1 Overview The fol

Página 7

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 96 Operating Instructions 6.2 Messages on the Detector Display E

Página 8

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 97 Message Remedial Action Bad flow cell p

Página 9 - 1 Introduction

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 98 Operating Instructions Message Remedial Action Filter control

Página 10 - UltiMate 3000 Series:

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 99 Message Remedial Action High dark signa

Página 11 - 1.2 Safety

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 100 Operating Instructions Message Remedial Action Lamp intensit

Página 12

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 101 Message Remedial Action Reference sign

Página 13

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 3 1.2 Safety The CE Mark label and cTUVus

Página 14

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 102 Operating Instructions Message Remedial Action VIS lamp sign

Página 15

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 103 When the detector is operated with Chro

Página 16

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 104 Operating Instructions 6.3 Chromeleon Diagnostics Messages I

Página 17

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 105 Test Result Remedial Action Zero order

Página 18

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 106 Operating Instructions Grating Motor Test failed Test Result

Página 19

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 107 Test Result Remedial Action Holmium li

Página 20

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 108 Operating Instructions Intensity Test failed In the test resu

Página 21 - 1.3 Intended Use

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 109 6.4 Operating Problems The following t

Página 22

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 110 Operating Instructions Problem Probable Cause Remedial Acti

Página 23

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 111 Problem Probable Cause Remedial Actio

Página 24

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 4 Operating Instructions At various points throughout the manual,

Página 25 - 2 Overview

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 112 Operating Instructions Problem Probable Cause Remedial Acti

Página 26 - 2.2 Operating Principle

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 113 Problem Probable Cause Remedial Actio

Página 27

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 114 Operating Instructions Problem Probable Cause Remedial Acti

Página 28 - 2.3 Configurations

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 115 7 Service 7.1 General Notes and Safet

Página 29

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 116 Operating Instructions If the original shipping container is

Página 30 - 2.4 Interior Components

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 117 7.2 Wavelength Calibration and Verific

Página 31 - 2.5 Front Panel Elements

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 118 Operating Instructions To perform wavelength verification in

Página 32 - 2.6 Rear Panel

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 119 7.3 Lamp Observe the following precaut

Página 33 - Fig. 5: DAC Board

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 120 Operating Instructions Tips: You can monitor the number of

Página 34

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 121 7.3.1 Replacing the Deuterium Lamp Des

Página 35 - 2.7 Fluid Connections

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 5 1.2.2 Safety Precautions When working wi

Página 36 - 2.8 Flow Cells

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 122 Operating Instructions 4. Loosen the two screws in the lamp

Página 37 - 2.9 Lamps

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 123 7.3.2 Installing or Replacing the Tung

Página 38 - 2.11 Detector Control

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 124 Operating Instructions When replacing the tungsten lamp Loo

Página 39

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 125 7.4 Flow Cell 7.4.1 Dummy Flow Cell T

Página 40

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 126 Operating Instructions 7.4.3 Replacing the Flow Cell Safety

Página 41 - 3 Installation

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 127 • Tubing connections between the adapt

Página 42 - Cardboard

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 128 Operating Instructions 7.4.3.1 Removing a Flow Cell (or Dumm

Página 43

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 129 7.4.3.3 Installing a Flow Cell (or Dum

Página 44

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 130 Operating Instructions 5. Remove the dummy plugs on the adap

Página 45 - 3.4 Connecting the Detector

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 131 7.5 Drying the Leak Sensor The leak se

Página 46

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 6 Operating Instructions • Hazardous gases Install the HPLC syst

Página 47 - USB Port

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 132 Operating Instructions 7.6 Replacing the Main Power Fuses

Página 48

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 133 7.7 Updating the Detector Firmware Th

Página 49

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 134 Operating Instructions 3. Right-click the detector in the ti

Página 50

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 135 8 Technical Information Optical design

Página 51

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 136 Operating Instructions Data collection rate: Single waveleng

Página 52 - Fig. 13: General page

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 137 Environmental conditions: Range of use:

Página 53 - Fig. 14: Detector page

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 138 Operating Instructions

Página 54

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 139 9 Accessories, Consumables, and Spare

Página 55

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 140 Operating Instructions 9.2 Optional Accessories Description

Página 56

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 141 Description Part No. Remarks Capillar

Página 57 - 4.1 Overview of Actions

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 7 • Hot surfaces ♦ During operation, the

Página 58

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 142 Operating Instructions 9.3 Consumables and Spare Parts The p

Página 59 - 4.3 Connecting Drain Tubing

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 143 Description Part No. Flow cell, analyt

Página 60

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 144 Operating Instructions Description Part No. Tungsten lamp 6

Página 61

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 145 10 Appendix 10.1 Common Mobile Phases

Página 62

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 146 Operating Instructions Functional Group Chromophore Wavelen

Página 63 - 5 Operation and Maintenance

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 147 10.2 Digital I/O The two digital I/O p

Página 64 - 5.1 Power-Up

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 148 Operating Instructions 10.3 Declaration of Conformity for Ho

Página 65 - Wavelength [nm]

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 149 11 Index A Absorbance Spectrum ...

Página 66 - 5.3 Detector Control

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 150 Operating Instructions F Firmware Download ...

Página 67

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 151 PCM-3000...

Página 68 - Fig. 18: Commands dialog box

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 8 Operating Instructions • Before switching from buffer to organ

Página 69

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 152 Operating Instructions

Página 70

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 9 1.2.3 Consignes de Sécurité Si vous util

Página 71

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 10 Operating Instructions • Gaz dangereux Installez le système H

Página 72

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 11 ♦ Portez des lunettes de protection lor

Página 73

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions

Página 74 - Fig. 23: Detector menus

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 12 Operating Instructions • Remplacez les cordons d'aliment

Página 75

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 13 1.3 Intended Use For Research Use Only.

Página 76

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 14 Operating Instructions If there is any question regarding appr

Página 77

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 15 1.4 Federal Communications Commission (

Página 78 - 5.4.2.4 Preferences Menu

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 16 Operating Instructions

Página 79

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 17 2 Overview 2.1 Unit Description The de

Página 80

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 18 Operating Instructions 2.2 Operating Principle Photometric de

Página 81

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 19 Fig. 1: Optics setup No. Component De

Página 82

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 20 Operating Instructions 2.3 Configurations The detector is ava

Página 83

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 21 In addition, the following is available

Página 84

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page I Declaration of Conformity (Original Decl

Página 85

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 22 Operating Instructions 2.4 Interior Components The front pane

Página 86

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 23 2.5 Front Panel Elements Fig. 3: Fron

Página 87

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 24 Operating Instructions 2.6 Rear Panel 1 2 5 7 3 4 6 Fig. 4:

Página 88

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 25 2.6.2 Fuse Cartridge The fuse cartridge

Página 89 -

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 26 Operating Instructions 2.6.6 pH and Conductivity Monitor (opt

Página 90

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 27 2.7 Fluid Connections The detector is d

Página 91

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 28 Operating Instructions 2.8 Flow Cells All flow cells (→ page

Página 92

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 29 2.9 Lamps The following lamps are avail

Página 93

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 30 Operating Instructions 2.10 Leak Sensor A leak sensor (→ Fig.

Página 94

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 31 Two modes of software control are availa

Página 95

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page II Operating Instructions

Página 96

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 32 Operating Instructions 2.12 System Wellness, Predictive Perfo

Página 97

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 33 3 Installation 3.1 Facility Requiremen

Página 98

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 34 Operating Instructions 2. Grasp the detector by the sides. Sl

Página 99

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 35 3.3 Positioning the Detector in the Ult

Página 100

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 36 Operating Instructions Fig. 10: Example for the rear panel co

Página 101

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 37 3.4 Connecting the Detector 3.4.1 Gene

Página 102

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 38 Operating Instructions 3.4.4 Connecting the Signal Cable (Dig

Página 103 - 6 Troubleshooting

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 39 4. Slide the left guide bar of the DAC

Página 104

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 40 Operating Instructions 3.4.5.3 Configuring the Analog Outputs

Página 105

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 41 3.5 Setting Up the Detector in Chromele

Página 106

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page i Table of Content 1 Introduction ...

Página 107

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 42 Operating Instructions 7. Windows Vista, Windows 7, and Windo

Página 108

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 43 3.5.2 Installing the Detector After th

Página 109

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 44 Operating Instructions 3.5.3 Configuring the Detector 3.5.3.1

Página 110

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 45 • Module Address Select the module addr

Página 111

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 46 Operating Instructions ♦ UV Lamp and VIS Lamp The UV Lamp che

Página 112

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 47 3.5.3.2 Changing the Configuration Prop

Página 113 - Filter Motor Test failed

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 48 Operating Instructions 3.6 Setting Up the Detector in DCMSLin

Página 114

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 49 4 Preparation for Operation (Startup)

Página 115

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 50 Operating Instructions 4.2 General Precautions for Connecting

Página 116

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 51 4.3 Connecting Drain Tubing To discharg

Página 117 - 6.4 Operating Problems

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page ii Operating Instructions 3.5.2 Installing the Detector ...

Página 118

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 52 Operating Instructions 4.4 Equilibrating the System Before us

Página 119

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 53 The equilibration panel shows the equili

Página 120

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 54 Operating Instructions

Página 121

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 55 5 Operation and Maintenance For informa

Página 122

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 56 Operating Instructions 5.1 Power-Up To start the detector for

Página 123 - 7 Service

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 57 5.2 Status Screens When the self test

Página 124

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 58 Operating Instructions 5.3 Detector Control 5.3.1 Operation

Página 125

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 59 5.3.2.1 Connecting to Chromeleon 1. St

Página 126

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 60 Operating Instructions 3. Press the F8 key or select Command

Página 127 - 7.3 Lamp

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 61 To open a control panel 1. On the View

Página 128

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page iii 7 Service ...

Página 129

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 62 Operating Instructions 5.3.2.3 Automated Control With automat

Página 130

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 63 2. Change the settings in the program a

Página 131

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 64 Operating Instructions 5.4 Display Screens (Function Keys and

Página 132

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 65 To … Select… Open the Main menu (→ page

Página 133 - 7.4 Flow Cell

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 66 Operating Instructions 5.4.2 Detector Menus Fig. 23 shows an

Página 134

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 67 5.4.2.1 General Menu Layout and Structu

Página 135

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 68 Operating Instructions To … Select … Toggle between two opera

Página 136

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 69 5.4.2.2 Main Menu The Main menu provide

Página 137 - Locating pin

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 70 Operating Instructions 5.4.2.4 Preferences Menu On the Prefe

Página 138

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 71 5.4.2.5 Diagnostics Menu The Diagnosti

Página 139 - 7.5 Drying the Leak Sensor

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page iv Operating Instructions

Página 140

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 72 Operating Instructions To … Select … Turn on the selected lam

Página 141

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 73 5.5 Information for Operating the Detec

Página 142

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 74 Operating Instructions 5.5.1.1 Wavelength When you select a w

Página 143 - 8 Technical Information

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 75 If the wavelengths are selected such tha

Página 144

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 76 Operating Instructions 5.5.1.2 Mobile Phases Mobile phase qua

Página 145

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 77 5.5.1.3 Mobile Phase Delivery System Th

Página 146

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 78 Operating Instructions 5.5.2 Setting the Wavelengths Set the

Página 147 - 9.1 Standard Accessories

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 79 To turn on the lamps manually • In Chro

Página 148 - 9.2 Optional Accessories

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 80 Operating Instructions To monitor the lamp intensity from the

Página 149

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 81 Monitoring the progress of data acquisit

Página 150

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 1 1 Introduction 1.1 How to Use this Manu

Página 151

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 82 Operating Instructions 5.5.6 Setting the Display Mode You can

Página 152

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 83 5.5.8 Recording the Temperature Signals

Página 153 - 10 Appendix

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 84 Operating Instructions 5.6 Special Functions in Chromeleon Th

Página 154

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 85 Fig. 26: UV Options page in the program

Página 155 - 10.2 Digital I/O

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 86 Operating Instructions • In multiple wavelength applications,

Página 156 - Holmium Oxide Glass Filter

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 87 5.6.1.4 Baseline Behavior If wavelength

Página 157 - 11 Index

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 88 Operating Instructions 5.6.2 Recording Absorbance Spectra The

Página 158

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 89 5.6.3 Predictive Performance Predictive

Página 159

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Page 90 Operating Instructions Color Description Orange (Only for mo

Página 160

UltiMate 3000 Series: VWD-3100 and VWD-3400RS Variable Wavelength Detectors Operating Instructions Page 91 5.6.5 Operational Qualification and Per

Comentários a estes Manuais

Sem comentários